Denver Instrument
5 Orville Dr.
Suite 200
Bohemia, NY 11716
800.321.1135
303.431.7255
303.423.4831 FAX
email
Links
Sitemap
|
Comprehensive Documentation.
A built-in printer provides data on every test. A unique header with initial and final weights, the program number and parameters, plus sample identification can be added to the ISO/GLP printout. Statistical analysis can be performed on up to 250 test results.
Great Choice for Multiple Users. Simply press the TEST key and the instrument will prompt the user through the analysis. Three features help with the uniformity of sample analysis. 1.) The target weight allows easy weigh-in of the sample and assures the correct sample size. 2.) Starting parameters for the test, including a time delay to be sure the instrument is ready. 3.) User ID tracks which operator performed the test while the printout provides a space for a signature.
Advanced Heating Element.
The IR-120 Moisture Analyzer features the new coil quartz infrared heating element, which is specially designed to provide the consistent heating you've come to expect from Denver Instrument. The heating element is housed in a motorized heating hood that can be opened at a keypress or set to close automatically before a test. The hood retracts into the body of the moisture analyzer providing direct, overhead access to the weighing pan for easy sample entry.
Features
- Quartz coil infrared heating element
- Single-block weighing technology with 0.001g resolution for precise
results
- Readability of 0.01%
- Internal balance calibration
- Store up to 100 custom programs
- Programmable one, two or three-step drying procedure
- Fully adjustable endpoint criteria to maximize efficiency and
accuracy
- Motorized heating hood for full access to sample
- Store up to 250 test results
- Built-in dot-matrix printer and RS232 port
- Security feature to prevent unauthorized changes
| IR-120 Moisture Analyzer Specifications |
| Working Range |
0.1 - 99.9% |
 |
| Measuring Method |
Infrared radiation, loss on drying |
 |
| Heat Source |
quartz coil infrared heater |
 |
| Temperature Setting |
40° - 210°C in one degree increments |
 |
| Temperature Control |
Platinum temperature sensor +/- 5°C |
 |
| Temperature Steps |
Programmable; one, two or three |
 |
| End of Analysis Mode |
Time-out, automatic slope or fully adjustable |
 |
| Units of Results |
% moisture, % solid, % moisture dry, weight loss (g or mg) |
 |
| Readability of Results |
0.01% |
 |
| Balance Capability |
120g |
 |
| Balance Readability |
1mg (0.001g) |
 |
| Program Storage |
100 programs |
 |
| Data Storage |
Statistics on 9,999 points |
 |
| Display |
Back-lit liquid crystal display |
 |
| Integrated Printer |
Dot matrix |
 |
| External I/O |
Serial unidirectional RS-232C |
 |
| Power requirements |
115 or 230 VAC, 50/60 Hz |
 |
External Dimensions (L x W x H) |
18.0 x 13.8 x 6.0" (457 x 351 x 152mm) |
 |
| Net Weight |
18 lbs. (8 kg) |
 |
IR-35 Moisture Analyzer
IR-35 Moisture Analyzer
IR-120 Moisture Analyzer
|
|